Potentiometer: comparing EMFs and finding internal resistance

Along a uniform wire carrying a steady current, V = kl. At the null point no current is drawn, so E₁/E₂ = l₁/l₂ and r = R(l₁/l₂ − 1).

PrincipleFor a steady current in a uniform wire, the potential drop is proportional to the length: V = kl
Potential gradientk = V/l, the potential drop per unit length of the wire, in V m⁻¹
Comparing EMFsE₁/E₂ = l₁/l₂, where l₁ and l₂ are the balance lengths
Internal resistancer = R(l₁/l₂ − 1); l₁ with the cell on open circuit, l₂ with R joined across it

Why a potentiometer?

A voltmeter must draw a little current to work. So across a cell it reads the terminal voltage, a little less than the EMF. A potentiometer measures a voltage without drawing any current, by a null method. It has a long uniform wire of manganin or nichrome, usually ten 1 m lengths (four in some laboratory models) joined by thick copper strips, beside a metre scale. A jockey makes contact at any point of the wire. A steady driver cell (an accumulator) and a rheostat send a steady current through the wire. Then the potential falls evenly along it: V = kl, where k is the potential gradient. The EMF of the driver cell must be more than any voltage to be measured, or no balance point is found.

How to draw the potentiometer circuits

  1. 1
    Main circuit: draw a long line AB for the wire. Join the + terminal of the driver cell to A, and its − terminal through a key K and a rheostat Rh to B.
  2. 2
    Comparing EMFs: join the + terminals of cells E₁ and E₂ to A. Join their − terminals to the two sides of a two-way key, and the middle of the key through a galvanometer G to the jockey J.
  3. 3
    Internal resistance: join the + terminal of cell E₁ to A, and its − terminal through G to the jockey. Put a resistance box R and a key K₁ directly across the cell.
  4. 4
    Mark the balance lengths l₁ and l₂ measured from A. Show the + terminals of all the cells joined to the same end, A.

Activity: comparing the EMFs of two cells

Aim: to compare the EMFs of a Leclanche cell (E₁) and a Daniell cell (E₂) with a potentiometer.

  1. 1
    Set up the main circuit and the two-way key circuit as drawn. Clean all wire ends with sandpaper.
  2. 2
    Close K and connect E₁. Touch the jockey at A and then at B: the galvanometer must deflect opposite ways. If not, raise the current with the rheostat.
  3. 3
    Slide the jockey to find the point of no deflection for E₁. Record the balance length l₁ from A.
  4. 4
    Without touching the rheostat, connect E₂ instead and find its balance length l₂.
  5. 5
    Change the rheostat setting and repeat twice more, recording l₁ and l₂ each time.
  6. 6★
    Result: E₁ = kl₁ and E₂ = kl₂, so E₁/E₂ = l₁/l₂. Take the mean of the ratios.

★ key step

Activity: internal resistance of a cell

Aim: to find the internal resistance r of a primary cell, such as a Leclanche cell, of EMF E₁.

  1. 1
    Set up the main circuit, and the cell circuit with a resistance box R and a key K₁ across the cell.
  2. 2
    With K₁ open, no current flows through the cell. Find the balance length l₁. Then E₁ = kl₁.
  3. 3
    Take out a small resistance R, say 5 Ω, and close K₁. The cell now drives a current through R, so its terminal voltage V falls.
  4. 4
    Quickly find the new balance length l₂, then open K₁. Then V = kl₂.
  5. 5
    Since E₁/V = (R + r)/R = l₁/l₂, the internal resistance is r = R(l₁/l₂ − 1).
  6. 6★
    Result: repeat for R = 6, 7, 8, 9 and 10 Ω. Observe: l₂ rises as R rises. Take the mean value of r.

★ key step

Worked example: three potentiometer numericals

Problem

(a) A cell of EMF 1.10 V balances at 330 cm. Another cell balances at 450 cm on the same wire. Find its EMF. (b) A cell balances at 300 cm on open circuit and at 250 cm when a 10 Ω resistor is joined across it. Find its internal resistance. (c) A 4 m wire carries a steady current with 2.0 V across it. Find the potential gradient.

Working
  1. (a) E₂/E₁ = l₂/l₁, so E₂ = 1.10 V × 450 ÷ 330 = 1.10 V × 1.364 = 1.50 V.
  2. (b) r = R(l₁/l₂ − 1) = 10 Ω × (300/250 − 1) = 10 Ω × 0.20 = 2.0 Ω.
  3. (c) k = V/l = 2.0 V ÷ 4 m = 0.50 V m⁻¹.
Answer

(a) 1.50 V; (b) 2.0 Ω; (c) 0.50 V m⁻¹

Potentiometer vs voltmeter

PotentiometerDraws no current at balance, so it measures the true EMF. It uses a null method, so the galvanometer need not be calibrated. A longer wire makes it more accurate.
VoltmeterMust draw a small current, so across a cell it reads the terminal voltage, a little less than the EMF. It is quick and handy, but less accurate.

A longer wire gives a smaller potential gradient, so each centimetre stands for a smaller voltage. That makes the potentiometer more sensitive.

In real life

The volume knob of a radio is a small potentiometer: a sliding contact on a resistance picks off part of the voltage. Laboratories use wire potentiometers to check the EMF of cells accurately, without drawing current from them.

Exam angle

1 marks: “What is the potential gradient of a potentiometer wire?” The potential drop per unit length of the wire, k = V/l, in V m⁻¹.

2 marks: “Why is a potentiometer preferred to a voltmeter for measuring the EMF of a cell?” At the balance point the potentiometer draws no current from the cell, so there is no Ir drop and it reads the true EMF. A voltmeter draws some current and reads a little less.

5 marks: “Explain the principle of a potentiometer. With a circuit diagram, describe how you would find the internal resistance of a cell with it.” Principle: V = kl for a steady current in a uniform wire. Balance at l₁ with the cell open (E = kl₁). Join R across the cell and balance at l₂ (V = kl₂). E/V = (R + r)/R = l₁/l₂, so r = R(l₁/l₂ − 1). Draw the main circuit and the cell circuit.

Remember it

No current, no lie: at the null point the cell gives no current, so the potentiometer reads its true EMF. Compare lengths, not meter readings.